SUBJECT: Small volume nebulization of medications to patients on mechanical ventilation.

PURPOSE: To define procedure for administration of nebulized medications to patients on mechanical ventilation.

POLICY:

1. Nebulized medications are given as per physician's order.

2. Nebulizers are removed from ventilator circuit and allowed to dry between treatments. Nebulizers will not be rinsed with tap water (they may be rinsed with sterile water or sterile normal saline).

3. Nebulizers are discarded and replaced with new daily and when visibly soiled.

INDICATIONS: Bronchodilator aerosol administration and evaluation of response are indicated whenever bronchoconstriction or increased airways resistance is documented or suspected in patients during mechanical ventilation. Some possible indications are:

  1. Known expiratory airflow obstruction.
  2. Previous demonstrated response to bronchodilator.
  3. Presence of auto-PEEP not eliminated with reduced rate, increased inspiratory flow, or decreased inspiratory to expiratory time ratio.
  4. Increased airway resistance as evidenced by increased peak inspiratory pressure and plateau pressure difference, wheezing or decreased breath sounds, intercostal and/or sternal retractions, or patient-ventilator dysynchrony.

CONTRAINDICATIONS:

EQUIPMENT:

PROCEDURE:

During mechanical ventilation, the deposition of drug to the lower respiratory tract is reduced. Standard doses may need to be adjusted to compensate for reduced delivery. Variables should be optimized to safely enhance medication delivery.

  1. Assess patient.
  2. Place medication in dry nebulizer.
  3. Place nebulizer securely in ventilator curcuit on inspiratory side aproximately 30 cm back from endotracheal tube. Assure no leaks are present.
  4. If heated humidifier is in use place it in pause mode during nebulization. If heat and moisture exchanger (HME) is in use remove it from the circuit during nebulization.
  5. Assure that a filter is in place in the expiratory limb between the patient and the ventilator.
  6. Optimize ventilator settings as patient tolerates to maximize medication deposition (Consider the following, if not otherwise contraindicated--(1) Use of a tidal volume > 500 mL for adults; (2) addition of an inspiratory pause or lower flows, which may improve pulmonary deposition of aerosol; however clinical judgment and patient evaluation must assure that the patient's inspiratory flow demands are met and risk of baro/volu-trauma is not increased (ie, the inspiratory-to-expiratory-time ratio is subjectively and physiologically appropriate, auto-PEEP is not increased, Pplat does not rise above 35 cmH2O); (3) because spontaneous breaths may improve aerosol delivery, spontaneous breathing should not be suppressed during aerosol therapy unless the patient's ability to trigger the ventilator is affected). Flow from the flowmeter to the nebulizer may affect the delivered tidal volume, the inspired oxygen concentration, and the patient's ability to trigger the ventilator. It may be necessary to decrease the set tidal volume. For a patient triggering the ventilator, the rate may need to be increased to maintain an appropriate minute ventilation
  7. Start flow to nebulizer at 6-10 LPM.
  8. Assess patient and ventilator. Make necessary modifications to safely optimize medication delivery.
  9. When medication has been delivered re-establish origional ventilator circuit configuration and ventilator settings.
  10. Assess patient.
  11. Perform patient/ventilator system check.

Documentation

PRECAUTIONS AND/OR POSSIBLE COMPLICATIONS:

INFECTION CONTROL:

SOURCES:

AARC Clinical Practice Guideline - Selection of Device, Administration of Bronchodilator, and Evaluation of Response to Therapy in Mechanically Ventilated Patients. Respir Care 1999;44(1):105-113